Title of article :
Extraction-flow injection spectrofluorimetric measurement of dissolved oxygen in environmental waters using 2-thionaphthol Original Research Article
Author/Authors :
Tadao Sakai، نويسنده , , Hisashi Takio، نويسنده , , Norio Teshima، نويسنده , , Harumitsu Nishikawa، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 2001
Pages :
5
From page :
117
To page :
121
Abstract :
A simple and highly sensitive flow injection analysis for the spectrofluorimetric determination of dissolved oxygen (DO) in environmental waters was developed. The decrease in fluorescence, which is based on the reaction of 2-thionaphthol with iodine liberated in Winkler’s method, was utilized as the DO detection system. The calibration graph was linear over the range 2×10−6–1.2×10−5 mol/l iodine. The relative standard deviation (n=8) was below 1% for the 6×10−6 mol/l iodine determination and the detection limit (S/N=3) was 4.9×10−7 mol/l. The sample frequency was 18 h−1.
Keywords :
Spectrofluorimetry , Extraction , 2-Thionaphthol , Dissolved oxygen , Flow injection
Journal title :
Analytica Chimica Acta
Serial Year :
2001
Journal title :
Analytica Chimica Acta
Record number :
1029767
Link To Document :
بازگشت