Title of article
Examination of interferences in flame atomic absorption spectrometry using a flow injection technique Original Research Article
Author/Authors
Pawel Koscielniak، نويسنده , , Marek Kozak، نويسنده ,
Issue Information
روزنامه با شماره پیاپی سال 2001
Pages
8
From page
187
To page
194
Abstract
Interference effects were examined in FAAS by means of specific measurement of flow injection peaks. In each given case, a single FI peak was examined — those parameters which define the inflexion points were measured. Parameters characterising a common region of two parts of a double FI peak were also taken into consideration. The interferences were studied using the example of tested samples containing calcium in the presence of aluminium and phosphates. It was revealed that the parameters considered can be measured with very good precision allowing one to detect even slight peak deformations caused by intereferences. Among the approaches applied, one using a double peak appeared to be most sensitive and reliable. It has been also noticed that the inflexion points of an FI peak may play a special role in the interference examinations.
Keywords
Interference effects , Flame atomic absorption spectrometry , Flow injection analysis
Journal title
Analytica Chimica Acta
Serial Year
2001
Journal title
Analytica Chimica Acta
Record number
1029776
Link To Document