Title of article
Analysis of solid samples by capillary electrophoresis using a gas extraction sampling device in a flow system Original Research Article
Author/Authors
Adriana G Lista، نويسنده , , Lourdes Arce، نويسنده , , Angel Rios، نويسنده , , Miguel Valc?rcel، نويسنده ,
Issue Information
روزنامه با شماره پیاپی سال 2001
Pages
8
From page
315
To page
322
Abstract
A gas extraction sampling device integrated in a continuous flow system (CFS) was coupled to a capillary electrophoresis equipment (CE) for the direct determination of trimethylamine (TMA) and related amines in fish sample. This arrangement allowed the direct introduction and treatment of these solid samples with a high level of automation. The amine compounds were extracted from fish samples in the sampling unit, then they were conditioned in the CFS and injected into the CE vials via a programmable arm. Methylamine, dimethylamine, trimethylamine, ethylamine, propylamine, diethylamine and triethylamine were separated and quantified. The concentration of trimethylamine (the major amine in fish) in the analysed real sample was ranged from 20 to 250 mg g−1.
Keywords
Automatic sample treatment , Capillary electrophoresis , fish , Amines
Journal title
Analytica Chimica Acta
Serial Year
2001
Journal title
Analytica Chimica Acta
Record number
1029791
Link To Document