• Title of article

    Analysis of solid samples by capillary electrophoresis using a gas extraction sampling device in a flow system Original Research Article

  • Author/Authors

    Adriana G Lista، نويسنده , , Lourdes Arce، نويسنده , , Angel Rios، نويسنده , , Miguel Valc?rcel، نويسنده ,

  • Issue Information
    روزنامه با شماره پیاپی سال 2001
  • Pages
    8
  • From page
    315
  • To page
    322
  • Abstract
    A gas extraction sampling device integrated in a continuous flow system (CFS) was coupled to a capillary electrophoresis equipment (CE) for the direct determination of trimethylamine (TMA) and related amines in fish sample. This arrangement allowed the direct introduction and treatment of these solid samples with a high level of automation. The amine compounds were extracted from fish samples in the sampling unit, then they were conditioned in the CFS and injected into the CE vials via a programmable arm. Methylamine, dimethylamine, trimethylamine, ethylamine, propylamine, diethylamine and triethylamine were separated and quantified. The concentration of trimethylamine (the major amine in fish) in the analysed real sample was ranged from 20 to 250 mg g−1.
  • Keywords
    Automatic sample treatment , Capillary electrophoresis , fish , Amines
  • Journal title
    Analytica Chimica Acta
  • Serial Year
    2001
  • Journal title
    Analytica Chimica Acta
  • Record number