Title of article :
Potentiometric determination of diazepam, bromazepam and clonazepam using solid contact ion-selective electrodes Original Research Article
Author/Authors :
A.A. Salem، نويسنده , , B.N. Barsoum، نويسنده , , E.L. Izake، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 2003
Abstract :
Newly developed solid contact ion-selective electrodes (SC-ISE) have been proposed for determining diazepam, bromazepam and clonazepam 1,4-benzodiazepines in pure forms and in pharmaceutical preparations. The electrodes are based on polyvinylchloride (PVC) membranes doped with drug-tetraphenyl borate (TPB) or drug-phosphotungestic acid (PTA) ion pair complexes as electroactive materials. The new electrodes proved highly selective with selectivity coefficients ranging from 10−4 to 10−6 and detection limits ranging from 0.1 to 0.63 μg/ml. Successful application of developed electrodes for drugs determinations in pharmaceutical preparations was obtained. Calibration graphs were linear over the ranges of 0.29–71.20, 0.79–31.62 and 0.32–31.57 μg/ml for diazepam, bromazepam and clonazepam, respectively with slopes of 50–65 mV per decade in the optimum pH range of 4–7. Relative average errors ≤1.72% and relative standard deviations ≤1.13% were obtained. Statistical student’s t-test and F-test showed insignificant systematic error between measured and real values and also between developed ion-selective electrodes methods and a standard HPLC method.
Keywords :
Diazepam , Bromazepam , Clonazepam , Pharmaceutical preparations , PVC , Benzodiazepines , Potentiometry , Ion-selective electrodes
Journal title :
Analytica Chimica Acta
Journal title :
Analytica Chimica Acta