• Title of article

    Flow injection-chemiluminescence determination of paraben preservative in food safety Original Research Article

  • Author/Authors

    Aung Myint Khaing، نويسنده , , Qunlin Zhang، نويسنده , , Lijuan Liu، نويسنده , , Hua Cui، نويسنده ,

  • Issue Information
    روزنامه با شماره پیاپی سال 2004
  • Pages
    6
  • From page
    119
  • To page
    124
  • Abstract
    The determination of a group of parabens including methylparaben, ethylparaben, propylparaben, and butylparaben using flow injection analysis technique with chemiluminescence detection has been carried out. The method is based on the enhancement by parabens of cerium(IV)-rhodamine 6G chemiluminescence reaction in sulfuric acid medium. Under the optimized conditions, the linear range of 4.5 × 10−10 to 1.5 × 10−7, 3.0 × 10−9 to 5.0 × 10−7, 5.0 × 10−9 to 1.0 × 10−6, and 8.0 × 10−9 to 8.0 × 10−7 g ml−1 and the detection limit of 3.4 × 10−10, 2.0 × 10−9, 2.0 × 10−9, and 5.0 × 10−9 g ml−1 were obtained for methylparaben, ethylparaben, propylparaben, and butylparaben, respectively. The relative standard deviations were less than 2.0%. The method shows higher sensitivity than most of the reported methods, which also require preconcentration or derivatization. It was successfully applied to the determination of ethylparaben in soy sauces without tedious pretreatment. The possible mechanism of this chemiluminescence reaction has been proposed due to that cerium(IV) was reduced by the products hydrolyzed by parabens to the excited-state cerium(III) in sulfuric acid medium and then energy was transferred from cerium(III)* to rhodamine 6G to form the excited-state rhodamine 6G, emitting light.
  • Keywords
    Flow injection , Paraben , Cerium(IV) , Rhodamine 6G , Chemiluminescence
  • Journal title
    Analytica Chimica Acta
  • Serial Year
    2004
  • Journal title
    Analytica Chimica Acta
  • Record number

    1030383