Title of article
Semi-permeable membrane device efficiency for sampling free soluble fraction of polycyclic aromatic hydrocarbons Original Research Article
Author/Authors
C. Miège، نويسنده , , C. Ravelet، نويسنده , , J.P. Croué، نويسنده , , J. Garric، نويسنده ,
Issue Information
روزنامه با شماره پیاپی سال 2005
Pages
8
From page
259
To page
266
Abstract
Semi-permeable membrane devices (SPMDs) have been used for KDOC (defined as the partition coefficient of hydrophobic organic contaminants between dissolved organic carbon and water) determination of five selected polycyclic aromatic hydrocarbons (PAHs) (fluoranthene, pyrene, benzo[b]fluoranthene, benzo[a]pyrene, benzo[e]pyrene). The Aldrich humic acid and the Suwannee river hydrophobic acid isolate (XAD8 isolate) were used as dissolved organic carbon. Uptake rate constants (ku and Rs) were evaluated and the SPMD method was calibrated based on a 3 h exposure time. For the five PAHs, the linearity was checked for water PAH concentration from 1 ng/L to 1 μg/L. The analytical detection limits ranged from 2 to 50 ng/L for the benzo[a]pyrene and the pyrene respectively. The proposed methodology is suitable for facile laboratory KDOC measurement and produces results similar to previously published values.
Keywords
KDOC , Semi-permeable membrane device (SPMD) , Polycyclic aromatic hydrocarbons (PAHs)
Journal title
Analytica Chimica Acta
Serial Year
2005
Journal title
Analytica Chimica Acta
Record number
1030619
Link To Document