Title of article :
Volatilization of in situ generated antimony trichloride using aqua regia vapors for the determination of ultra trace impurities in high-purity antimony using inductively coupled plasma-mass spectrometry Original Research Article
Author/Authors :
A.C. Sahayam، نويسنده , , Shiuh-Jen Jiang، نويسنده , , Chia-Ching Wan، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 2007
Pages :
5
From page :
214
To page :
218
Abstract :
Volatilization of Sb has been carried out using vapors generated from aqua regia for the determination of non-volatile impurities present in antimony. The antimony trichloride (SbCl3) thus generated was volatilized >99.9% at 250 °C in 90 min. Recoveries of spiked impurities Al, Mn, Co, Ni, Cu, Sr, Ag, Cd, Ba, Pb were found to be >90% (47% for Pb) after volatilization of the matrix element. The only reagent used (aqua regia vapors) was generated in situ in a closed environment minimizing the external contamination. The procedure offers experimental blanks and method detection limits at lower ng g−1 using ICP-MS. The method has been applied for the determination of impurities in two Sb samples of 99.5 and 99.999% (5N) purity. The developed method was validated using bromide volatilization procedure. The results obtained by the present method were also compared with those of Sector Field ICP-MS (SF-ICP-MS).
Keywords :
High-purity materials , Antimony , Antimony trichloride , Volatilization , Inductively coupled plasma-mass spectrometry , Aqua regia vapors
Journal title :
Analytica Chimica Acta
Serial Year :
2007
Journal title :
Analytica Chimica Acta
Record number :
1031121