Title of article :
Flow-injection determination of acetone with diazotized anthranilic acid through a fluorescent reaction intermediate Original Research Article
Author/Authors :
C. Garc?a de Mar?a، نويسنده , , K.B. Hueso Dom?nguez، نويسنده , , N. Mart?n Garrido، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 2007
Abstract :
Acetone and diazotized anthranilic acid react in alkaline solution, giving a fluorescent intermediate that can be measured at excitation and emission wavelengths of 305 and 395 nm, respectively. Based on this, a fluorimetric flow-injection method is proposed for the determination of acetone in aqueous solution. Under the proposed conditions, acetone can be detected at concentrations higher than 8 × 10−7 M, with a linear application range from 1 × 10−6 to 2 × 10−4 M and an R.S.D. of 2.7% (1.0 × 10−5 M, n = 10). A sampling frequency of 24 h−1 is achieved. Some potentially interfering species are investigated.
Keywords :
Acetone , Flow injection , Fluorimetry , Anthranilic acid
Journal title :