Title of article :
Flow injection visible diffuse reflectance quantitative analysis of nickel Original Research Article
Author/Authors :
Matthieu Tubino، نويسنده , , Carlos A.R. Queiroz، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 2007
Pages :
6
From page :
199
To page :
204
Abstract :
Flow injection (FI) methodology, using diffuse reflectance in the visible region of the spectrum, for the analysis of nickel, precipitated in the form of dimethylglyoximate, is presented. A reflectance cell, constructed in polytetrafluoroethylene, using a LED (light emitting diode) as light source and a LDR (light dependent resistor) as detector, is described. The analytical signal (S) correlates with nickel concentration (C) between 1.6 × 10−4 and 6.6 × 10−4 mol L−1. This correlation is described by the equation S = −1.108 + 3.314 × 104C − 2.081 × 107C2 (r = 0.9996). The experimentally observed limit of detection is about 1.3 × 10−4 mol L−1, as in lower concentrations the formation of precipitate is not observed. The experimental quantitation limit is about 1.6 × 10−4 mol L−1. The mean R.S.D. (relative standard deviation) is about 2.7%. Samples containing nickel were analyzed and the results obtained in this method were compared with those of other methods using the statistical Studentʹs t-test.
Keywords :
Dimethylglyoxime , Visible , Flow injection , Determination , nickel , Diffuse reflectance
Journal title :
Analytica Chimica Acta
Serial Year :
2007
Journal title :
Analytica Chimica Acta
Record number :
1031201
Link To Document :
بازگشت