• Title of article

    Automatic determination of fat in milk by use of a flow injection system with a piezoelectric detector Original Research Article

  • Author/Authors

    Lisbeth Manganiello، نويسنده , , Angel Rios، نويسنده , , Miguel Valcarcel، نويسنده , , Angel Ligero، نويسنده , , Teresa Tena، نويسنده ,

  • Issue Information
    روزنامه با شماره پیاپی سال 2000
  • Pages
    7
  • From page
    309
  • To page
    315
  • Abstract
    A new method for the direct determination of total fat in milk by use of a quartz crystal microbalance was developed. The method is based on a (micro) extraction procedure that is carried out in a flow injection system. Samples are diluted with an ethanol–water mixture and injected into the flow system, where they are mixed with n-hexane in an extraction coil. The analyte is determined in the organic phase. The calibration thus achieved is linear in the range 0.20–0.45% w/v, and the relative standard deviation is ±3.2% (n = 11; P = 0.05). The throughput is 12 samples per hour. The proposed method was used to determine fat matter in milk samples; the results were found to be competitive with those of official methods for the same purpose.
  • Keywords
    Flow system , Piezoelectric detection , Fat , Milk
  • Journal title
    Analytica Chimica Acta
  • Serial Year
    2000
  • Journal title
    Analytica Chimica Acta
  • Record number