Title of article :
Polymeric membrane silver-ion selective electrodes based on bis(dialkyldithiophosphates) Original Research Article
Author/Authors :
Dong Liu، نويسنده , , Jianhong Liu، نويسنده , , Deyu Tian، نويسنده , , Weiliang Hong، نويسنده , , Xiaomin Zhou، نويسنده , , Jimmy C. Yu، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 2000
Abstract :
Poly(vinyl chloride) (PVC) membrane electrodes based on bis(dialkyldithiophosphates) carriers were prepared and tested for selective detection of silver ions. A suitable lipophilicity of the carrier and appropriate co-ordination ability were found to be essential for designing an electrode with good response characteristics. A PVC membrane with 5% bis(diethyldithiophosphates) ionophore and 65% o-NPOE plasticizer satisfies these requirements. This electrode exhibits linear response over a concentration range of 10−1–10−6 mol l−1 Ag+ with a slope of 57.3 mV/dec. This is a fast and simple analytical method, and it has been applied successfully in the determination of silver ions in electroplating wastewater.
Keywords :
PVC membrane , Bis(dialkyldithiophosphates) , Ionophores , Silver-ion selective electrode
Journal title :
Analytica Chimica Acta
Journal title :
Analytica Chimica Acta