Measurement of parts per million levels of potassium hydroxide in polyether polyol streams Original Research Article
Author/Authors :
Sofia Galanis، نويسنده , , Purnendu K. Dasgupta، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 2001
Pages :
10
From page :
101
To page :
110
Abstract :
An automated reverse flow injection (FI) based continuous analyzer for KOH in polyols is presented. A solution of bromocresol green (BCG) in 2-propanol, with or without added mineral acid, is injected into the sample polyol stream in a single-line FI manifold maintained at 110°C (dispersion factor of 25 and 50 for a dilute aqueous stream at room temperature with 30 and 10 μl of solution injected) and the blue indicator dianion formed upon reaction with KOH is measured at 605 nm by a light emitting diode (LED) — fiber optic based absorbance detector. Injections of 10 μl of 10 mM BCG and 30 μl of 15 mM BCG, 50 mM HCl were respectively chosen to cover desired ranges of 0–20 and 20–120 mg KOH/kg polyol (ppm KOH). The absorbance signal is linear with KOH concentration (r2=0.9949 in both cases). Measurements could be made in the present system every 4 min. Deliberate alteration of water content of the polyol stream in the range of interest was found to have no discernible effect on the measurement.
Keywords :
Polyether polyols , KOH , Light emitting diode , Flow injection analysis