• Title of article

    Measurements of buffer capacities of reducing gases against oxygen using a zirconia oxygen pump-gauge Original Research Article

  • Author/Authors

    Hitoshi Taimatsu، نويسنده , , Hiroyuki Kaneko، نويسنده ,

  • Issue Information
    روزنامه با شماره پیاپی سال 2001
  • Pages
    8
  • From page
    247
  • To page
    254
  • Abstract
    A small amount of a reducing gas, H2, CO or CH4, in Ar was titrated with oxygen over a wide range of oxygen partial pressures using a zirconia oxygen pump-gauge in a similar way to the pH titration in aqueous solutions. Concentrations of the reducing gases in Ar were determined from the equivalence points in the titration curves. The buffer capacities against oxygen were theoretically formulated for the mixtures, and were compared with those analyzed from the measured titration curves. Precise values of the buffer capacity were obtained for the reducing gases above a total oxygen demand of 100 ppm, and the measured buffer capacities deviated from theoretical ones at much lower reducing gas concentrations. Several factors influencing the measurement of the buffer capacity and the limit of the measurement were discussed.
  • Keywords
    Buffer capacity , Gas titration , Reducing gas , Zirconia oxygen pump , Zirconia oxygen sensor
  • Journal title
    Analytica Chimica Acta
  • Serial Year
    2001
  • Journal title
    Analytica Chimica Acta
  • Record number