• Title of article

    Potentiometric ion-selective silicon sensors for the on-line monitoring of blood electrolytes Original Research Article

  • Author/Authors

    Scheipers، A. نويسنده , , O Wa?mus، نويسنده , , C Sundermeier، نويسنده , , Eshold، J. نويسنده , , Th Wei?، نويسنده , , M Gitter، نويسنده , , B Ro?، نويسنده , , M Knoll، نويسنده ,

  • Issue Information
    روزنامه با شماره پیاپی سال 2001
  • Pages
    10
  • From page
    29
  • To page
    38
  • Abstract
    A non-invasive on-line monitoring system for blood electrolytes has been developed on the basis of potentiometric ion-selective silicon sensors. The microsensors were fabricated in the patented containment technology owned by ICB. Additionally, multianalyte sensors with two and three containment sensor elements integrated on chip were realized. The new polysiloxane membrane functionalized with trifluoropropyl groups with improved adhesion on the silicon oxide surface were used as matrix membrane for all prepared solid state microsensors. The following containment sensors have been developed: potassium, calcium, sodium, and pH. They were characterized in physiological test solutions. The excellent reliability of the potassium sensors has been demonstrated successfully in clinical trials with real circulating whole blood samples and on extracorporal animal blood circulation. First simultaneous measurements of potassium, sodium and pH have also been performed in vitro using human blood samples.
  • Keywords
    Potentiometry , Silicon technology , Blood electrolytes , On-line monitoring , Containment technology
  • Journal title
    Analytica Chimica Acta
  • Serial Year
    2001
  • Journal title
    Analytica Chimica Acta
  • Record number

    1032482