Title of article :
Chemiluminescence detection of paracetamol by a luminol-permanganate based reaction Original Research Article
Author/Authors :
D Easwaramoorthy، نويسنده , , Yueh-Chuan Yu، نويسنده , , Hsuan-Jung Huang، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 2001
Pages :
6
From page :
95
To page :
100
Abstract :
The flow injection analysis of paracetamol was developed based on its inhibitory effect on a luminol-permanganate chemiluminescence system. With a system containing solutions of 2.5 mM luminol, 5.0 μM KMnO4 and pH 12.0 phosphate buffer (K2HPO4/NaOH), a detection limit of 1.0×10−8 M (1.5 ppb) and a linear dynamic range of 2.5×10−8 to 2.5×10−7 M can be obtained. A repetitive injection of 1.0×10−7 M paracetamol solutions gave satisfactory reproducibility with a relative standard deviation of 2.3% (n=12). Possible interferences from major excipients in pharmaceuticals and other related compounds were also studied. Contents of paracetamol in two locally available pharmaceuticals were analyzed. From the recovery of 98.2–104.4% obtained, the feasibility of the developed method for real sample analysis was confirmed.
Keywords :
Luminol-permanganate , Chemiluminescence , Inhibition effect , Paracetamol
Journal title :
Analytica Chimica Acta
Serial Year :
2001
Journal title :
Analytica Chimica Acta
Record number :
1032490
Link To Document :
بازگشت