Title of article :
Separation of niobium(V) and tantalum(V) by capillary electrophoresis with chemiluminescence detection Original Research Article
Author/Authors :
Erbao Liu، نويسنده , , Yan-Ming Liu، نويسنده , , Jieke Cheng، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 2001
Abstract :
A new method for simultaneous determination of niobium(V) and tantalum(V) is described. Capillary electrophoresis (CE) with chemiluminescence (CL) detection was used in the separation and detection of niobium(V) and tantalum(V) which catalyze the CL reaction of luminol with hydrogen peroxide (H2O2). The mixing mode used in this work allowed the CL reagents (luminol and H2O2) and the analytes to meet simultaneously at the detection window while in conventional ion chromatography chemiluminescence (ICCL) the CL reagents are mixed in advance. Therefore, in this work the luminol as a component of the separation carrier electrolyte avoided the loss of light signal which occurs when luminol and hydrogen peroxide are mixed in advance in ICCL. Signal enhancement was achieved by sample stacking injection in the electrophoresis process. A sample volume of 20 nl was injected. The metal ions were separated in less than 7 min with sodium dihydrogen phosphate buffer (pH 2.5, 10 mmol/l). The detection limits of this method for niobium(V) and tantalum(V) (S/N=3) are 826 zmol (4.13×10−10 mol/l) and 640 amol (3.2×10−8 mol/l), respectively. The sensitivity of the metal ions was satisfactory.
Keywords :
Chemiluminescence , Electrophoresis , Niobium , Tantalum
Journal title :
Analytica Chimica Acta
Journal title :
Analytica Chimica Acta