Title of article :
All-solid-state planar miniature ion-selective chloride electrode Original Research Article
Author/Authors :
R. Zieli?ska، نويسنده , , E. Mulik، نويسنده , , A. Michalska-Jednoralska، نويسنده , , S. Achmatowicz، نويسنده , , M. Maj-?urawska، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 2002
Abstract :
All-solid-state ion-selective electrodes with plastic membrane (poly(vinyl chloride) (PVC), bis(2-ethylhexyl) sebacate (DOS), methyltri-n-tetradecylammonium chloride (MTTACl)), a conducting poly(pyrrole) (PPy) film doped either with chloride ions (PPyCl) or hexacyanoferrate(II) ions (PPyFeCN), and glassy carbon (GC) or screen-printed graphite layer (S-PG) as an inner electric contact were investigated. All the electrodes show close to Nernstian response, but their lifetimes vary. The at least 2-months lifetime of screen-printed electrodes is only achieved for the electrodes containing PPyFeCN (cation-exchanging film). Shorter lifetime of other screen-printed electrodes, i.e. without PPy, or with PPyCl (anion-exchanging film), was attributed to the diffusion of anionic products of the hydrolysis of organic components of the graphite paste used to prepare the electric contact. The properties of miniature, screen-printed electrodes comprising PPyFeCN solid contact, were comparable to those ion-selective electrodes with PPy solid contact (regardless the ion-exchanging characteristic of the polymer) deposited on GC electric contact.
Keywords :
All-solid-state potentiometric sensor , Chloride sensor , Screen-printed electrodes , Poly(pyrrole)
Journal title :
Analytica Chimica Acta
Journal title :
Analytica Chimica Acta