Title of article :
Immunochromatographic flow-injection method for detection of human serum immunoglobulin G antibodies to Helicobacter pylori Original Research Article
Author/Authors :
D Narinesingh، نويسنده , , T.T Ngo، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 2002
Abstract :
This study reports on the development of an immunochromatographic flow-injection (FI) method for the quantitation of human serum IgG antibodies to Helicobacter pylori. Patients’ sera were injected into the carrier stream of a FI manifold incorporating an H. pylori immuno-adsorbent reactor. The immuno-adsorbent was prepared by covalently linking H. pylori antigens to periodate oxidized Sepharose CL-4B. Any H. pylori antibody present in the serum is bound to the immuno-sorbent. The bound antibody was quantified by injecting into the carrier stream anti-human IgG peroxidase conjugate followed by TMB/H2O2 as the enzyme substrate. The FI system was optimized with respect to antigen loading (1.0 mg/ml gel), bioreactor length (3.0 cm), enzyme conjugate (0.2 mg/ml) and sample loop size (250 μl). One hundred clinical samples were analyzed for their H. pylori antibody concentrations and the results evaluated with respect to their receiver–operator characteristics (ROC). When a cut-off absorbance of between 0.7 and 0.75 was used for positive and negative samples a specificity (>95%), a sensitivity (>90%)and an overall accuracy (>94%) were obtained.
Keywords :
Flow-injection analysis , Immunochromatography , H. pylori antibodies
Journal title :
Analytica Chimica Acta
Journal title :
Analytica Chimica Acta