Title of article :
Development of integrated chemiluminescence flow sensor for the determination of adrenaline and isoprenaline Original Research Article
Author/Authors :
Guojun Zhou، نويسنده , , Guofang Zhang، نويسنده , , Hongyuan Chen، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 2002
Pages :
7
From page :
257
To page :
263
Abstract :
A novel integrated chemiluminescence (CL) flow sensor for the determination of adrenaline and isoprenaline is developed based on the enhancing effect of analytes on CL emission of luminol oxidized by periodate in alkaline solution. The analytical reagents luminol and periodate are immobilized on anion exchange resins, respectively, and packed in a glass tube to construct a reagentless sensor. The proposed sensor allows the determination of adrenaline and isoprenaline over the range from 2.0×10−8 to 1.0×10−5 g ml−1 and 2.0×10−7 to 5.0×10−5 g ml−1, respectively. The detection limits are 7.0×10−9 g ml−1 for adrenaline and 5.0×10−8 g ml−1 for isoprenaline with a relative standard deviation of 1.7% for the 1.0×10−7 g ml−1 adrenaline (n=11) and 2.1% for 1.0×10−6 g ml−1 isoprenaline (n=11). The sample throughput was 60 samples h−1. The sensor has been successfully applied to the determination of adrenaline and isoprenaline in pharmaceutical preparations.
Keywords :
Sensor , Adrenaline , Chemiluminescence , Isoprenaline , Flow injection analysis
Journal title :
Analytica Chimica Acta
Serial Year :
2002
Journal title :
Analytica Chimica Acta
Record number :
1033112
Link To Document :
بازگشت