Title of article
In situ osmotic analyzer for the year-long continuous determination of Fe in hydrothermal systems Original Research Article
Author/Authors
Thomas P. Chapin، نويسنده , , Hans W. Jannasch، نويسنده , , Kenneth S. Johnson، نويسنده ,
Issue Information
روزنامه با شماره پیاپی سال 2002
Pages
10
From page
265
To page
274
Abstract
A submersible osmotically pumped analyzer (Fe-OsmoAnalyzer) has been adapted for the long-term continuous high resolution monitoring of iron(II) and (III) in deep-sea hydrothermal vents. Ferrozine is used as the colorimetric reagent. The detection limit for our analyzer is 0.1 μM for a 0.7 cm pathlength with a linear response up to 50 μM. The Fe-OsmoAnalyzer uses periodic injections of standards and blanks to self calibrate and can operate unattended for over a year collecting data every 15 min. The Fe-OsmoAnalyzer was field tested during a 6 week test deployment at 900 m in Monterey Bay and a year-long deployment at 1100 m off the coast of Hawaii. A second year-long deployment in a low temperature hydrothermal vent on the Juan de Fuca Ridge provided high resolution continuous monitoring of Fe and was able to distinguish tidally influenced fluctuations in Fe concentrations.
Keywords
In situ , Seawater , Long-term , Hydrothermal , Osmotic pumps , Iron determination
Journal title
Analytica Chimica Acta
Serial Year
2002
Journal title
Analytica Chimica Acta
Record number
1033113