Title of article :
Flow injection analysis applied to a voltammetric electronic tongue Original Research Article
Author/Authors :
F Winquist، نويسنده , , E Rydberg، نويسنده , , S Holmin، نويسنده , , C Krantz-Rülcker، نويسنده , , I Lundstr?m، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 2002
Pages :
14
From page :
159
To page :
172
Abstract :
A measurement system, based on flow injection analysis (FIA) technique applied to a voltammetric electronic tongue is described. A reference solution was thus continuously pumped through a cell with a voltammetric electronic tongue, and test samples were injected into the flow stream. Responses were obtained by measuring the resulting pulse height. The FIA technique offered several advantages, since relative measurements are performed, the system is less influenced by sensor baseline drift, calibration samples and/or washing solutions can be injected within a measurement series, and the system is well adapted for automatization. The system was used to analyze standard solutions of H2O2, KCl, CuNO3, K4[Fe(CN)6], K3[Fe(CN)6] and NaCl, and results obtained were treated with multivariate data analysis. Principal component analysis performed showed that electrode drift could be considerably decreased, and the set-up was also used for classification of different apple juices.
Keywords :
Electronic tongue , Taste sensor , Flow injection analysis , Pulsed voltammetry , Principal component analysis
Journal title :
Analytica Chimica Acta
Serial Year :
2002
Journal title :
Analytica Chimica Acta
Record number :
1033295
Link To Document :
بازگشت