Title of article :
Development of a method for estimating an accurate equivalence point in nickel titration of cyanide ions Original Research Article
Author/Authors :
Toshihiro Suzuki، نويسنده , , Akiharu Hioki، نويسنده , , Masayasu Kurahashi، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 2003
Abstract :
In a nickel titration of cyanide ions using murexide as indicator, an accurate equivalence point was determined by a non-linear least-squares curve-fitting for a titration curve. This method was developed to establish a standard solution for cyanide ions. In a curve-fitting procedure, a theoretical titration curve was calculated, assuming that nickel ion formed only a 1:4 Ni2+:CN− complex with cyanide ions and formed only a 1:1 complex with murexide. Results of the curve-fitting were reasonable at any pH and any indicator concentration studied. The combined standard uncertainty for a concentration of a 1000 mg kg−1 cyanide solution by this method was 0.079%.
Keywords :
Complexometric titration , Equivalence point , Murexide , Uncertainty , Nickel ion , Cyanide ion
Journal title :
Analytica Chimica Acta
Journal title :
Analytica Chimica Acta