• Title of article

    Amperometric phenol biosensor based on sol–gel silicate/Nafion composite film Original Research Article

  • Author/Authors

    Min Ah Kim، نويسنده , , Won-Yong Lee، نويسنده ,

  • Issue Information
    روزنامه با شماره پیاپی سال 2003
  • Pages
    8
  • From page
    143
  • To page
    150
  • Abstract
    An amperometric biosensor based on tyrosinase immobilized in silicate/Nafion composite film has been developed for the determination of phenolic compounds. The Nafion polymer in the composite was used not only to overcome the brittleness of the pure sol–gel-derived silicate film but also to increase the long-term stability of the biosensor. Tyrosinase was immobilized by a thin film of silicate/Nafion composite on a glassy carbon electrode. Phenolic compounds were determined by the direct reduction of biocatalytically-liberated quinone species at −200 mV versus Ag/AgCl (3 M NaCl). The process parameters for the fabrication of the enzyme electrode and various experimental variables such as pH and operating potential were explored for optimum analytical performance of the enzyme electrode. The biosensor can reach 95% of steady-state current in about 15 s. The sensitivities of the biosensor for catechol and phenol were 200 and 46 mA/M, respectively. A detection limit of 0.35 mM catechol was obtained with a signal-to-noise ratio of 3. The enzyme electrode retained 74% of its initial activity after 2 weeks of storage in 50 mM phosphate buffer at pH 7.
  • Keywords
    Tyrosinase , Phenolic compounds , Silicate/Nafion composite film , Sol–gel technique , Amperometric biosensor
  • Journal title
    Analytica Chimica Acta
  • Serial Year
    2003
  • Journal title
    Analytica Chimica Acta
  • Record number