• Title of article

    Coated wire ion-selective electrode for palladium(II) and its application to catalyst analysis Original Research Article

  • Author/Authors

    Mohammad Aghamohammadi، نويسنده , , Naader Alizadeh، نويسنده ,

  • Issue Information
    روزنامه با شماره پیاپی سال 2003
  • Pages
    8
  • From page
    299
  • To page
    306
  • Abstract
    The construction, performance and application of polymeric membrane (PME) and coated wire (CWE) palladium(II)-selective electrodes based on the ion pair between tetra bromopalladate(II) and hexadecylpyridinium cation (HDP+) in a poly(vinyl chloride) matrix, plasticized with o-NPOE are described. The influence of membrane composition, bromide ion concentration and pH on the potentiometric responses of electrodes were investigated. Nernstian responses were obtained for the two type of electrodes with low limits of detection (1.0×10−6 M for PME and 5.0×10−8 M for CWE). The potentiometric responses are independent of the pH of test solution in the range 3–8. The response time of electrodes are fast (30 s for PME and 10 s for CWE), and they can be used for at least 3 months without any considerable divergence in potentials. The proposed electrodes revealed good selectivity for palladium(II) respect to different cations and anions. They were used to the direct potentiometric determination of palladium(II) in silicon-alumina catalysts.
  • Keywords
    Sodium , 12-Crown-3 , Bis(crown ether)s , Ion-selective electrode , Conformational analysis , Computational chemistry
  • Journal title
    Analytica Chimica Acta
  • Serial Year
    2003
  • Journal title
    Analytica Chimica Acta
  • Record number

    1033433