• Title of article

    Hydride generation atomic fluorescence spectrometric determination of ultratraces of selenium and tellurium in cow milk Original Research Article

  • Author/Authors

    Patricia Cava-Montesinos، نويسنده , , M.Luisa Cervera، نويسنده , , Agustin Pastor، نويسنده , , Miguel de la Guardia، نويسنده ,

  • Issue Information
    روزنامه با شماره پیاپی سال 2003
  • Pages
    10
  • From page
    291
  • To page
    300
  • Abstract
    A sensitive procedure has been developed for selenium and tellurium determination in milk by hydride generation atomic fluorescence spectrometry (HG-AFS) after microwave-assisted sample digestion. The method provides sensitivity values of 1591 and 997 fluorescence units ng−1 ml−1 with detection limits of 0.005 and 0.015 ng ml−1 for Se and Te, respectively. The application of the developed methodology to the analysis of cow milk samples of the Spanish market evidenced the presence of concentration ranges from 11.1 to 26.0 ng ml−1 for Se, and from 1.04 to 9.7 ng ml−1 for Te having found a good comparability with data obtained after dry-ashing of samples.
  • Keywords
    Selenium determination , Tellurium determination , Milk analysis , Hydride generation atomic fluorescence , Microwave-assisted digestion
  • Journal title
    Analytica Chimica Acta
  • Serial Year
    2003
  • Journal title
    Analytica Chimica Acta
  • Record number