Title of article :
Liquid core waveguide-based optical spectrometry for field estimation of dissolved BTEX compounds in groundwater: A feasibility study Original Research Article
Author/Authors :
Henrik Larsson، نويسنده , , Purnendu K. Dasgupta، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 2003
Abstract :
The utility of a liquid core waveguide (LCW) system for acting as a sentry monitor for compounds such as benzene, toluene, ethylbenzene, xylene (BTEX) in water was examined. A vapor-permeable LCW suitable for long path length absorbance spectroscopy based on a Teflon AF 2400 tube was fabricated. Multiwavelength spectroscopy in the near-UV was carried out using a fiber optic-based flashlamp-photodiode array (PDA) combination with hexane as the solvent in the waveguide core. Using multicomponent calibration, quantitation of benzene and toluene accurate to 6±5% could be conducted at sub-mg l−1 levels in mixtures after a sampling period of 10 min.
Keywords :
Liquid core waveguide , Absorbance spectroscopy , BTEX compounds , Teflon AF
Journal title :
Analytica Chimica Acta
Journal title :
Analytica Chimica Acta