• Title of article

    Analysis of hydroxyurea in human plasma by high performance liquid chromatography with electrochemical detection Original Research Article

  • Author/Authors

    Y.J. Jong، نويسنده , , Tsuen-Ho Hsu، نويسنده , , H.L. Wu، نويسنده , , H.S. Kou، نويسنده , , S.M. Wu، نويسنده ,

  • Issue Information
    روزنامه با شماره پیاپی سال 2003
  • Pages
    8
  • From page
    223
  • To page
    230
  • Abstract
    An improved HPLC-electrochemical detection (ED) method is described for the analysis of hydroxyurea (HU) in human plasma. After extraction process, HU was determined on a C18 column (250 mm×4 mm) by the mobile phase (25 mM sodium acetate–acetonitrile, 97.5:2.5; pH 6.5). The regression equations were linear (r>0.9990). The precision and accuracy of intra- and inter-batches were all below 5% for relative standard deviation (R.S.D.) and relative error (R.E.). Based on 20 μl of plasma, the limit of detection was 0.3 μM for HU (S/N=3, injection 10 μl). This method was applied for the HU drug monitoring of patients with myelofibrosis or polycythemia vera.
  • Keywords
    HPLC-ED , Patients , hydroxyurea
  • Journal title
    Analytica Chimica Acta
  • Serial Year
    2003
  • Journal title
    Analytica Chimica Acta
  • Record number