Title of article :
Flow injection preconcentration by chemofiltration and spectrophotometric determination of Gd Original Research Article
Author/Authors :
Irma E De Vito، نويسنده , , Roberto A. Olsina، نويسنده , , Julio Raba، نويسنده , , Adriana N. Masi، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 2004
Abstract :
A new on-line Gd preconcentration and determination system associated to flow injection (FI) method was developed. 2,2′-(1,8-dihydroxy-3,6-disulfonaphthylene-2,7-bisazo) bisbenzenearsonic acid (Arsenazo III) was used as a complexing agent at pH 2.5. A reactor containing the polyamide membrane was used for the retention of the Gd complex by chemofiltration. The complex was then removed from the reactor with buffer solution pH 9. The variables affecting the combined on-line preconcentration–absorptiometric method have been evaluated and optimised. The coupling of the on-line preconcentration and spectrophotometric flow through detection led to a detection limit of 15 μg l−1 for a preconcentration time of 5 min at 2 ml/min. The method was successfully applied to biological samples.
Keywords :
Flow injection , Absorptiometric method , Gadolinium in urine sample , Preconcentration and determination
Journal title :
Analytica Chimica Acta
Journal title :
Analytica Chimica Acta