Title of article :
Droplet-based microfluidic lab-on-a-chip for glucose detection Original Research Article
Author/Authors :
Vijay Srinivasan، نويسنده , , Advanced Liquid Logic Vamsee K. Pamula، نويسنده , , Richard B. Fair، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 2004
Pages :
6
From page :
145
To page :
150
Abstract :
A microfluidic lab-on-a-chip (LoC) platform for in vitro measurement of glucose for clinical diagnostic applications is presented in this paper. The LoC uses a discrete droplet format in contrast to conventional continuous flow microfluidic systems. The droplets act as solution-phase reaction chambers and are manipulated using the electrowetting effect. Glucose is measured using a colorimetric enzyme-kinetic method based on Trinder’s reaction. The color change is detected using an absorbance measurement system consisting of a light emitting diode and a photodiode. The linear range of the assay is 9–100 mg/dl using a sample dilution factor of 2 and 15–300 mg/dl using a sample dilution factor of 3. The results obtained on the electrowetting system compare favorably with conventional measurements done on a spectrophotometer, indicating that there is no change in enzyme activity under electrowetting conditions.
Keywords :
Lab-on-a-chip , Droplet , Microfluidics , Glucose , Electrowetting