• Title of article

    On-line elimination of spectral interference of iron matrix in the flame atomic absorption determination of zinc by anion-exchange separation Original Research Article

  • Author/Authors

    Jamal A. Sweileh، نويسنده , , Eman M. El-Nemma، نويسنده ,

  • Issue Information
    روزنامه با شماره پیاپی سال 2004
  • Pages
    6
  • From page
    287
  • To page
    292
  • Abstract
    A flow injection analysis (FIA) system has been developed for the flame atomic absorption spectroscopic (FAAS) determination of zinc in iron matrix. The spectral line interference of iron at 213.859 nm was eliminated by on-line separation using a micro-column of strong anion-exchange resin (Dowex 1-X8). The zinc chloro complexes were retained from 2 M HCl solution while most of the iron chloro complexes were passed to waste. For a 2% iron solution, matrix removal efficiency was 98.2% which means that positive spectral line interference of iron at the Zn line was reduced from 0.42 to 0.008 μg ml−1 Zn. The optimized flow injection system can handle up to 48 samples with good precision (less than 3.5% relative standard deviation (R.S.D.)) in the working range of 0.075–2.2 μg ml−1 Zn. Comparative analysis of a certified reference material and synthetic sample solutions containing traces of Zn in 2% Fe by the proposed method and by graphite furnace atomic absorption spectroscopy (GFAAS) showed no evidence of analytical bias at the 95% confidence level.
  • Keywords
    Flow injection analysis , Atomic absorption , Spectral interference , Iron , Ion-exchange , zinc
  • Journal title
    Analytica Chimica Acta
  • Serial Year
    2004
  • Journal title
    Analytica Chimica Acta
  • Record number

    1034362