Title of article
Simultaneous on-line preconcentration and determination of trace metals in environmental samples by flow injection combined with inductively coupled plasma mass spectrometry using a nanometer-sized alumina packed micro-column Original Research Article
Author/Authors
Jun Yin، نويسنده , , Zucheng Jiang، نويسنده , , Gang Chang، نويسنده , , Bin Hu، نويسنده ,
Issue Information
روزنامه با شماره پیاپی سال 2005
Pages
7
From page
333
To page
339
Abstract
A flow injection (FI) on-line preconcentration procedure by using a nanometer-sized alumina packed micro-column coupled to inductively coupled plasma mass spectrometry (ICP-MS) was described for simultaneous determination of trace metals (V, Cr, Mn, Co, Ni, Cu, Zn, Cd and Pb) in the environmental samples. The effects of pH value, sample flow rate, preconcentration time, and interfering ions on the preconcentration of analytes have been investigated. Under the optimized operating conditions, the adsorption capacity of the nanometer-sized alumina for V, Cr, Mn, Co, Ni, Cu, Zn, Cd and Pb were found to be 11.7, 13.6, 15.7, 9.5, 12.2, 13.3, 17.1, 17.7 and 17.5 mg g−1, respectively. With 60 s preconcentration time and 60 s elution time, an enrichment factor of 5 and the sampling frequency of 15 h−1 were obtained. The proposed method has been applied to the determination of trace metals in environmental certified reference materials and natural water samples with satisfactory results.
Keywords
Nanometer-sized alumina , Flow injection , On-line preconcentration , Metal ions , ICP-MS , Environmental samples
Journal title
Analytica Chimica Acta
Serial Year
2005
Journal title
Analytica Chimica Acta
Record number
1034693
Link To Document