    Enzyme coated glass pH-electrode: Its fabrication and applications in the determination of urea in blood samples Original Research Article

  • Author/Authors

    Rachana Sahney، نويسنده , , B.K. Puri، نويسنده , , S. Anand، نويسنده ,

  • Issue Information
    روزنامه با شماره پیاپی سال 2005
  • Pages
    5
  • From page
    157
  • To page
    161
  • Abstract
    A new enzyme coated electrode for the determination of urea in blood samples has been developed. It is based on the encapsulation of urease enzyme in the porous silicate matrix by the sol–gel technique on a glass electrode for the purpose of sensing urea in blood samples. Various parameters like the effect of pH, selection of a suitable buffer of appropriate concentration and interference of common substances in blood samples have been evaluated to optimize the conditions for the determination of urea. The electrode can be used for the determination of urea in the concentration range 0.03–30.0 mM in a solution. The detection limit of the present enzyme-coated electrode is found to be 52 μg/ml of urea. The relative standard deviation for the electrode-to-electrode reproducibility is found to be 2.4% for the determination of 0.1 mM of urea (six replicate electrodes). Sol–gel matrix containing immobilized enzyme was stable for about 25 days at ∼4 °C with 80% urease activity. Urea content in various clinical blood samples has been estimated using this electrode and the results are found to be in good agreement with the standard clinical methods as reported in the literature.
  • Keywords
    Enzyme-coated electrode , Urea sensor , Sol–gel technique , Urea determination , Blood samples
  • Journal title
    Analytica Chimica Acta
  • Serial Year
    2005
  • Journal title
    Analytica Chimica Acta
  • Record number

    1034725