Title of article :
Characteristics of immersion angle dependence of the resonant frequency shift of the quartz crystal microbalance in solutions Original Research Article
Author/Authors :
Minoru Yoshimoto، نويسنده , , Shun-ichiro Yoshimura، نويسنده , , Shigeru Kurosawa، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 2006
Abstract :
We investigated the immersion angle dependence of the resonant frequency shift (ΔF) of the quartz crystal microbalance (QCM) in a Newtonian liquid from the point of view of the QCM oscillation direction. In this paper, the three cases of immersion angles were employed. In , the angles of both the surface with the electrode (x–z surface) and the QCM oscillation direction varied at the same time. In , the QCM oscillation direction changed while the x–z surface of the QCM was kept parallel to the gravity direction. In , the angle of the x–z surface of the QCM changed, while the QCM oscillation direction was constantly kept perpendicular to the gravity direction. This study elucidated that the ΔF values changed with the immersion angles in , and unchanged in . As a result, we have found that only when the x–z surface of the QCM is not parallel to the gravity direction, the immersion angle dependence of ΔF appears. Namely, it has become clear that the immersion angle dependence of ΔF has no relation to the QCM oscillation direction.
Keywords :
Oscillation direction , Immersion angle , QCM
Journal title :
Analytica Chimica Acta
Journal title :
Analytica Chimica Acta