Title of article :
Biosensor based on platinum chips for glucose determination Original Research Article
Author/Authors :
M. Asunci?n Alonso Lomillo، نويسنده , , Javier Gonzalo Ruiz، نويسنده , , F. Javier Mu?oz Pascual، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 2005
Pages :
6
From page :
209
To page :
214
Abstract :
Amperometric glucose biosensors based on a Pt microelectrode chip are described. Horseradish peroxidase (HRP) and glucose oxidase (GOX) were co-immobilized on a platinum microelectrode by electropolymerization. Two different compounds, namely ferrocyanide (Fe) and meldolaʹs blue (MB), were used as mediators to transfer electrons between the enzymes and the electrode. Experimental variables were optimised using experimental design methodology, by a 23 central composite design, taking the intensity registered for a given glucose solution as response. Given the reproducibility (residual standard deviation (R.S.D.), 9.122% with n = 3) and repeatability (R.S.D., 7.095% with n = 4) reached by the Fe-based microbiosensor under the optimum conditions, the analysis of glucose in grape juice, 319.46 g l−1 ± 37.12 (n = 3 and α = 0.05), was successfully accomplished.
Keywords :
Microsystem technologies , Microbiosensors , Electropolymerization , experimental design , Glucose
Journal title :
Analytica Chimica Acta
Serial Year :
2005
Journal title :
Analytica Chimica Acta
Record number :
1035530
Link To Document :
بازگشت