• Title of article

    Increasing robustness and sensitivity of protein microarrays through microagitation and automation Original Research Article

  • Author/Authors

    M. Hartmann، نويسنده , , A. Toegl، نويسنده , , R. Kirchner، نويسنده , , M.F. Templin، نويسنده , , T.O. Joos، نويسنده ,

  • Issue Information
    روزنامه با شماره پیاپی سال 2006
  • Pages
    8
  • From page
    66
  • To page
    73
  • Abstract
    Assay systems that employ protein microarrays for the analysis of complex samples are powerful tools to generate a high amount of data from a limiting amount of sample. Due to miniaturization, these systems are susceptible to fluctuations during signal generation and the use of uniform conditions for sample incubation and during the assay procedure is required to get reproducible results. Diffusion limits may prevent constant conditions on all parts of the array and can lead to the decease of the sensitivity of the array. Therefore, we set-up an automated assay system integrating a novel microagitation device using surface acoustic wave (SAW) technology. Multiplexed assays for the detection of autoantibodies from human serum and sandwich immunoassay for the detection of matrix metalloproteases (MMPs) were used to evaluate the system. Diffusion-rate limited solid phase reactions were enhanced by microagitation using the SAW technology resulting in up to three-fold higher signals.
  • Keywords
    Automation , Surface acoustic waves (SAW) , Microagitation , Protein microarrays , Diffusion limitation
  • Journal title
    Analytica Chimica Acta
  • Serial Year
    2006
  • Journal title
    Analytica Chimica Acta
  • Record number