    Phase-resolved detection in ion-mobility spectrometry Original Research Article

  • Author/Authors

    Andrew W. Szumlas، نويسنده , , Gary M. Hieftje، نويسنده ,

  • Issue Information
    روزنامه با شماره پیاپی سال 2006
  • Pages
    10
  • From page
    45
  • To page
    54
  • Abstract
    Throughput has been improved in an ion-mobility spectrometer by means of continuous ion-beam modulation and phase-resolved detection. Conventional ion-mobility spectrometry (IMS) utilizes brief packets of ions for separation in the gas phase. In contrast, this preliminary report outlines a new detection method based on phase resolution and involves monitoring the ion signal continuously, which raises the duty cycle of the instrument from approximately 1% (conventional IMS) to 50%. The greater duty cycle is shown to improve signal-to-noise ratios (S/N). However, the higher S/N levels did not translate into higher sensitivity values when calibration plots were constructed. It is demonstrated that even analytes with small differences in drift times can be modulated at frequencies that maximize the measured phase differences between ions. Phase-resolved signals can be processed with a Fourier transform and chemometric data reduction to simplify identification of the analyte from its phase spectra.
  • Keywords
    Enhanced resolution , Higher sensitivity , Ion-mobility spectrometry , Phase resolution , Fourier transform