• Title of article

    Multi-pumping flow system for the determination of dissolved orthophosphate and dissolved organic phosphorus in wastewater samples Original Research Article

  • Author/Authors

    Carmen Pons، نويسنده , , Ildik? V. T?th، نويسنده , , Antonio O.S.S. Rangel، نويسنده , , Rafael Forteza، نويسنده , , Victor Cerda، نويسنده ,

  • Issue Information
    روزنامه با شماره پیاپی سال 2006
  • Pages
    7
  • From page
    148
  • To page
    154
  • Abstract
    A multi-pumping flow system (MPFS) for the spectrophotometric determination of dissolved orthophosphate and dissolved organic phosphorus in wastewater samples is proposed. The determination of orthophosphate is based on the vanadomolybdate method. In-line ultraviolet photo-oxidation is employed to mineralise organic phosphorus to orthophosphate prior to detection. A solenoid valve allows the deviation of the flow towards the UV-lamp to carry out the determination of organic phosphorus. Calibration was found to be linear up to 20 mg P L−1, with a detection limit (3sb/S) of 0.08 mg P L−1, an injection throughput of 75 injections h−1 and a repeatability (R.S.D.) of 0.6% for the direct determination of orthophosphate. On the other hand, calibration graphs were linear up to 40 mg P L−1, with a detection limit (3sb/S) of 0.5 mg P L−1, an injection throughput of 11 injections h−1 and a repeatability (R.S.D.) inferior to 2.3% for the procedures involving UV photo-oxidation.
  • Keywords
    UV photo-oxidation , Multi-pumping flow system , Orthophosphate , Organic phosphorus
  • Journal title
    Analytica Chimica Acta
  • Serial Year
    2006
  • Journal title
    Analytica Chimica Acta
  • Record number

    1036004