Title of article
Electrogenerated chemiluminescence biosensor based on Ru(bpy)32+ and dehydrogenase immobilized in sol–gel/chitosan/poly(sodium 4-styrene sulfonate) composite material Original Research Article
Author/Authors
Lihua Zhang، نويسنده , , Zhiai Xu، نويسنده , , Shaojun Dong، نويسنده ,
Issue Information
روزنامه با شماره پیاپی سال 2006
Pages
5
From page
52
To page
56
Abstract
A new electrogenerated chemiluminescence biosensor was fabricated by immobilizing ECL reagent Ru(bpy)32+ and alcohol dehydrogenase in sol–gel/chitosan/poly(sodium 4-styrene sulfonate) (PSS) organically modified composite material. The component PSS was used to immobilize ECL reagent Ru(bpy)32+ by ion-exchange, while the addition of chitosan was to prevent the cracking of conventional sol–gel-derived glasses and provide biocompatible microenvironment for alcohol dehydrogenase. Such biosensor combined enzymatic selectivity with the sensitivity of ECL detection for quantification of enzyme substrate and it was much simpler than previous double-layer design. The detection limit was 9.3 × 10−6 M for alcohol (S/N = 3) with a linear range from 2.79 × 10−5 to 5.78 × 10−2 M. With ECL detection, the biosensor exhibited wide linear range, high sensitivity and good stability.
Keywords
Electrogenerated chemiluminescence (ECL) , Ru(bpy)32+ , Alcohol dehydrogenase , Sol–gel/chitosan/PSS , Biosensor