Title of article :
Method optimization for the determination of four mercury species by micro-liquid chromatography–inductively coupled plasma mass spectrometry coupling in environmental water samples Original Research Article
Author/Authors :
A. Castillo، نويسنده , , A.F. Roig-Navarro، نويسنده , , O.J. Pozo، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 2006
Abstract :
A method based on the coupling microHPLC–microneb-ICPMS has been developed for Hg(II), MeHg+, EtHg+ and PhHg+ species. Gradient elution using methanol and l-cysteine at pH 3.0 allowed the chromatographic separation of all species in less than 13 min (total analysis time 15 min). The direct coupling of microLC to ICPMS through a Micromist nebulizer permits the analysis of environmental water without sample pretreatment and derivatization steps. Nebulizer type, organic modifier and column length were the main parameters tested. The methanol content and pH of the mobile phase greatly affected the retention time and sensitivity of the method. Key factors to obtain high signal to noise ratio, at concentrations below 1 μg L−1, were found to be the nebulization step and traces of Hg present in the complexing agent. A detailed optimization of carrier and make up gas flow rates have enabled the nebulization of the methanol gradient elution with good mass transport efficiency, low organic solvent loading into the plasma and excellent precision.
Keywords :
Mercury speciation , Microbore column , Micro-liquid chromatography , Micronebulizer , Water , Inductively coupled plasma mass spectrometry