Title of article
Determination of gas phase triacetone triperoxide with aspiration ion mobility spectrometry and gas chromatography–mass spectrometry Original Research Article
Author/Authors
Riikka-Marjaana R?s?nen، نويسنده , , Marjaana Nousiainen، نويسنده , , Kaleva Per?korpi، نويسنده , , Mika Sillanp??، نويسنده , , Lauri Polari، نويسنده , , Osmo Anttalainen، نويسنده , , Mikko Utriainen، نويسنده ,
Issue Information
روزنامه با شماره پیاپی سال 2008
Pages
7
From page
59
To page
65
Abstract
Aspiration ion mobility spectrometry (IMS) has been used for the first time to screen 3,3,6,6,9,9-hexamethyl-1,2,4,5,7,8-hexaoxacyclononane explosive, the most commonly known as triacetone triperoxide (TATP). Gaseous TATP was generated from synthesized solid compound, sublimed and directed to a portable chemical detection system comprised of an aspiration-type IMS detector and six semiconductor sensors. Different unknown TATP gas phase concentrations were produced and corresponding IMS and semiconductor responses were measured. The experimental concentrations were determined by gas chromatography–mass spectrometry (GC–MS). The results evidenced that the monitored compound in the gas phase was TATP. In addition, the determined TATP concentrations and corresponding IMS intensities showed that the IMS response values were proportional to the measured TATP concentrations.
Keywords
Semiconductor sensor , Aspiration ion mobility spectrometry , Triacetone triperoxide , Improvised explosives
Journal title
Analytica Chimica Acta
Serial Year
2008
Journal title
Analytica Chimica Acta
Record number
1036196
Link To Document