• Title of article

    Determination of selected polychlorinated biphenyls in soil by miniaturised ultrasonic solvent extraction and gas chromatography-mass-selective detection Original Research Article

  • Author/Authors

    Mehmet Emin Aydin، نويسنده , , Ali Tor، نويسنده , , Senar Ozcan، نويسنده ,

  • Issue Information
    روزنامه با شماره پیاپی سال 2006
  • Pages
    6
  • From page
    232
  • To page
    237
  • Abstract
    Miniaturised ultrasonic solvent extraction procedure was developed for the determination of selected polychlorinated biphenyls (PCBs) in soil samples by gas chromatography-mass-selective detection by using 23 factorial experimental design. Recoveries of PCBs from fortified soil samples are over 90% for three different fortification levels between 40 and 120 μg kg−1, and relative standard deviations of the recoveries are below 7%. The limits of detection (LODs) ranged from 0.003 to 0.006 μg kg−1. The performance of the proposed method was compared to traditional shake flask extraction method on the spiked real soil sample and extraction methods showed comparable efficiencies. Proposed miniaturised ultrasonic solvent extraction offers several advantages, i.e., reducing sample requirement for measurement of target compound, less solvent consumption and reducing the costs associated with solvent purchase and waste disposal.
  • Keywords
    Micro-scale extraction , Polychlorinated biphenyls , Soil , factorial design , Ultrasonic solvent extraction
  • Journal title
    Analytica Chimica Acta
  • Serial Year
    2006
  • Journal title
    Analytica Chimica Acta
  • Record number