Use of hydrogen peroxide to achieve interference-free stripping voltammetric determination of copper at the bismuth-film electrode Original Research Article
Author/Authors
Wagner F. Pacheco، نويسنده , , Eliane M. Miguel، نويسنده , , Gabriel V. Ramos، نويسنده , , Carlos E. Cardoso، نويسنده , , Percio A.M. Farias، نويسنده , , Ricardo Q. Aucelio، نويسنده ,
Issue Information
روزنامه با شماره پیاپی سال 2008
Pages
6
From page
22
To page
27
Abstract
In this work, a new approach is presented to allow interference-free determination of Cu (II) by stripping voltammetry using the bismuth-film electrode. The addition of hydrogen peroxide to the electroanalytical cell has promoted complete resolution between re-dissolution peaks of Bi (III) and Cu (II). The absence of interference could be evaluated by the correlation coefficient (r > 0.99) between Cu (II) concentration and its shifted current peak (at +212 mV) while achieving a slightly fluctuation of the bismuth current peak at −180 mV. Studies were performed aiming towards the optimum conditions for trace determination of Cu (II) using hydrogen peroxide. The methodology was applied to a real sample (sugarcane spirits) and the results were compared to those from graphite furnace atomic absorption spectrometry. The analytical parameters of merit and the results of the analysis indicated that the analytical methodology could be readily used for trace determination of Cu (II).