Title of article :
Selectivity enhancement of anion-responsive electrodes by pulsed chronopotentiometry Original Research Article
Author/Authors :
Kebede L. Gemene، نويسنده , , Alexey Shvarev، نويسنده , , Eric Bakker، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 2007
Abstract :
A large and robust selectivity improvement of ion-selective electrodes is presented for the measurement of abundant ions. An improvement in selectivity by more than two orders of magnitude has been attained for the hydrophilic chloride ions measured in a dilute background of the lipophilic ions perchlorate and salicylate in a pulsed chronopotentiometric measurement mode. This is attributed to a robust kinetic discrimination of the dilute lipophilic ions in this measuring mode, which is not possible to achieve in classical potentiometry. Maximum tolerable concentrations of the interfering ions are found to be on the order of 30 μM before causing substantial changes in potential. As an example of practical relevance, the robust detection of chloride in 72 μM salicylate (reflecting 1:10 diluted blood) with a detection limit of 0.5 mM chloride is demonstrated. Corresponding potentiometric sensors did not give a useful chloride response under these conditions.
Keywords :
Ion-selective electrodes , Selectivity enhancement , Clinical analysis , Anion detection , Solvent polymeric membrane
Journal title :
Analytica Chimica Acta
Journal title :
Analytica Chimica Acta