    Dynamic method as a simple approach for wide range pH measurements using optodes Original Research Article

  • Author/Authors

    A. Safavi، نويسنده , , A.R. Banazadeh، نويسنده ,

  • Issue Information
    روزنامه با شماره پیاپی سال 2007
  • Pages
    6
  • From page
    326
  • To page
    331
  • Abstract
    In this paper, a flow system equipped with an optode has been suggested for wide range pH measurements. Triacetyl cellulose was used as the optode membrane in which different pH indicators were immobilized. For extending the pH range, the dynamic response rather than the steady-state response of the optode was measured. Since diffusion is the main process governing the system response, different parameters having influence on diffusion of the analyte into the membrane were optimized. Under the optimum conditions, wide range pH determination (up to 11 pH units) is simply achieved regardless of the pKa of the pH indicator immobilized in the membrane. To validate the application of the method different indicators with different structures and pKa values were tested and the results were all confirming the precision and accuracy of the method. The suggested method also has combined advantages of flow systems together with inherent advantages of kinetic systems.
  • Keywords
    Optode , Wide range pH determination , Dynamic
  • Journal title
    Analytica Chimica Acta
  • Serial Year
    2007
  • Journal title
    Analytica Chimica Acta
  • Record number