Title of article
Anodic stripping voltammetry of antimony using gold nanoparticle-modified carbon screen-printed electrodes Original Research Article
Author/Authors
Olga Dom?nguez Renedo، نويسنده , , M. Julia Arcos Mart?nez، نويسنده ,
Issue Information
روزنامه با شماره پیاپی سال 2007
Pages
6
From page
255
To page
260
Abstract
Carbon screen-printed electrodes (CSPE) modified with gold nanoparticles present an interesting alternative in the determination of antimony using differential pulse anodic stripping voltammetry. Metallic gold nanoparticles deposits have been obtained by direct electrochemical deposition. Scanning electron microscopy measurements show that the electrochemically synthesized gold nanoparticles are deposited in aggregated form. Any undue effects caused by the presence of foreign ions in the solution were also analyzed to ensure that common interferents in the determination of antimony by ASV. The detection limit for Sb(III) obtained was 9.44 × 10−10 M. In terms of reproducibility, the precision of the above mentioned method in %R.S.D. values was calculated at 2.69% (n = 10). The method was applied to determine levels of antimony in seawater samples and pharmaceutical preparations.
Keywords
Gold nanoparticles , Screen-printed electrodes , Anodic stripping voltammetry , Antimony
Journal title
Analytica Chimica Acta
Serial Year
2007
Journal title
Analytica Chimica Acta
Record number
1037154
Link To Document