Title of article :
Simplified paired-emitter–detector-diodes-based photometry with improved sensitivity Original Research Article
Author/Authors :
?ukasz Tymecki، نويسنده , , Robert Koncki، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 2009
Pages :
5
From page :
73
To page :
77
Abstract :
Paired emitter–detector diodes (PEDDs) coupled with potentiometric pH-meter or multimeter have been characterized as complete instruments for common photometric measurements as well as optical flow-through detectors. The analytical characteristics of investigated devices have been illustrated by the use of two paired red LEDs for the determination of bromothymol blue (BTB) as a model analyte. The developed devices allow rapid and sensitive BTB detection in the ppm range of concentration. Measurements with pH-meter resulted in a wide linearity of PEDD response, whereas PEDDs coupled with ordinary voltmeter offer a significant enhancement of sensitivity. Some practical applications of developed detector for the analytical and bioanalytical measurements have been indicated.
Keywords :
Photometry , Optical detector , Light-emitting diode , Flow analysis
Journal title :
Analytica Chimica Acta
Serial Year :
2009
Journal title :
Analytica Chimica Acta
Record number :
1037248
Link To Document :
بازگشت