Title of article :
Metalloporphyrin-based acetate-selective electrodes as detectors for enzymatic acetylcholine determination in flow-injection analysis system Original Research Article
Author/Authors :
Lukasz Gorski، نويسنده , , Monika Mroczkiewicz، نويسنده , , Mariusz Pietrzak، نويسنده , , Elzbieta Malinowska، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 2009
Abstract :
In this work two Zr(IV)-porphyrins were tested as potential acetate-selective ionophores. It is shown that these compounds show increased selectivity towards acetate ion, with selectivity sequence: Cl− < NO3− < SCN− < ClO4− < acetate− < F−, when incorporated into polymeric ion-selective membranes. Among tested ionophores, Zr(IV)-tetra(tert-butylphenyl)porphyrin was found to be the best in terms of response time (20 s) and lower detection limit (2 × 10−4 M acetate−).
Designed electrodes were used as detectors in flow-injection analysis system for acetylcholine determination. Enzymatic hydrolysis was used to generate acetate ions from acetylcholine, employing acetylcholinesterase immobilized on Amberlite. The optimization of enzymatic reactions and flow-injection analysis system configuration were performed and enzyme immobilization procedures have been evaluated. It is shown that resulting flow-injection system exhibits good working parameters, such as reproducibility (standard deviation ≤3.5%), sampling rate (60 samples per hour) and lifetime (over a week).
Keywords :
Flow-injection analysis , Potentiometry , Acetate , Polymeric-membrane ion-selective electrodes , Acetylcholine
Journal title :
Analytica Chimica Acta