Abstract :
A simple, low cost and portable microfluidic system based on a two-point alkaline picrate kinetic reaction has been developed for the determination of urinary creatinine. The creatinine reacts with picric acid under alkaline conditions, forming an orange-red colour, which is monitored on PDMS microchip using a portable miniature fibre optic spectrometer at 510 nm. A linear range was displayed from 0 to 40 mg L−1 creatinine (r2 = 0.997) with a detection limit of 3.3 mg L−1 (S/N = 3). On-chip absorbance signals are reproducible, with relative standard deviations (RSDs) of 7.1%, when evaluated with 20 mg L−1 creatinine (n = 10). The standard curves in which the intra-run CVs (4.7–6.8%) and inter-run CVs (7.9%) obtained were performed on three different days and exhibited good reproducibility. The method was highly correlated with the conventional spectrophotometric method when real urine samples were evaluated (r2 = 0.948; n = 15).
Keywords :
Creatinine determination , Jaffé reaction , Microfluidic system , Poly(dimethylsiloxane) , On-chip absorbance detection