Title of article
Micro-Winkler titration method for dissolved oxygen concentration measurement Original Research Article
Author/Authors
Irja Helm، نويسنده , , Lauri Jalukse، نويسنده , , Martin Vilbaste، نويسنده , , Ivo Leito، نويسنده ,
Issue Information
روزنامه با شماره پیاپی سال 2009
Pages
7
From page
167
To page
173
Abstract
In this report a gravimetric micro-Winkler titration method for determination of dissolved oxygen concentration in water is presented. Mathematical model of the method taking into account all influence factors is derived and an uncertainty analysis is carried out to determine the uncertainty contributions of all influence factors. The method is highly accurate: the relative expanded uncertainties (k = 2) are around 1% in the case of small (9–10 g) water samples. The uncertainty analysis carried out in characterizing the uncertainty of the method is the most comprehensive published for a micro-Winkler method, resulting in experimentally obtained estimates for all uncertainty sources of practical significance (around 20 uncertainty sources altogether).
Keywords
Measurement uncertainty , Uncertainty budget , Gravimetric titration , Micro-Winkler titration , Dissolved oxygen , Primary method
Journal title
Analytica Chimica Acta
Serial Year
2009
Journal title
Analytica Chimica Acta
Record number
1037439