• Title of article

    Direct determination of melamine in dairy products by gas chromatography/mass spectrometry with coupled column separation Original Research Article

  • Author/Authors

    Xiao-min Xu، نويسنده , , Yiping Ren، نويسنده , , Yun Zhu، نويسنده , , Zengxuan Cai، نويسنده , , Jianlong Han، نويسنده , , Baifen Huang، نويسنده , , Yan Zhu، نويسنده ,

  • Issue Information
    روزنامه با شماره پیاپی سال 2009
  • Pages
    5
  • From page
    39
  • To page
    43
  • Abstract
    A coupled capillary column system was developed for the qualitative and quantitative determination of melamine with isotope internal standard in dairy products by gas chromatography/mass spectrometry (GC/MS) without derivatization. A 30 m of DB-5ms ((5%-phenyl)-methylpolysiloxane, 0.25 mm i.d., 0.25 μm df) coupled with a 1.5 m of Innowax (polyethylene glycol, 0.32 mm i.d., 0.25 μm df) by a quartz capillary column connector was introduced as separation column. Three advantages were discussed for the coupled system. The sample was fortified with a ring-labeled 13C315N3-melamine as an isotope internal standard and extracted by 1% of trichloroacetic acid aqueous solution. 2.2% of lead acetate solution was then added to deposit protein in the sample matrix. After purification by cation exchange cartridge, the sample solution was directly injected and detected by GC/MS. A six-point calibration curve ranging from 0.05 to 2 mg kg−1 of melamine in sample was used to establish instrument response. The recovery was 93.9–102% with relative standard deviation from 3.1 to 8.7% when isotope internal standard used. The calculated method detection limit was 0.01 mg kg−1.
  • Keywords
    Melamine , Coupled column , Direct determination , Dairy products , Gas chromatography/mass spectrometry (GC/MS) , Isotope internal standard
  • Journal title
    Analytica Chimica Acta
  • Serial Year
    2009
  • Journal title
    Analytica Chimica Acta
  • Record number

    1037495